Brookeld Engineering Labs., Inc. Page 71 Manual No. M13-2100
V.4 Admin Functions
TheAdminFunctionsmenuincludessettingsrelatedtoaccessandbasicinstrumentmanagement
oftheDV3T.TheAdminFunctionsmenuisonlyavailabletotheAdministratorwhenLogInis
required.UsersandPowerUsersdonothaveaccessrightstotheAdminFunctions.Ninesettings
are available including:
Log In and Lock Out - RequireUserIDandPassword.DeneLockOutparameters.
Users and Access - DeneUserID,Password,Usercapabilities.AddUser.
SetTimeandDate - Setthetimeanddate.
Backup and Import - UpdateInstrumentFirmware,addLanguagele,BackUp
data and settings.
DefaultPath - DenethelogicfortheDefaultPathforsavingdataandtests.
SettingsReset - ResetbasicsettingstoDV3Tfactorydefault.
DeviceReset - ResettheentireDV3Ttofactorydefault.
CalibrationReminder - Establishuseandfrequencyofcalibrationreminder.
SaveAuditTrail - ExporttheAudittrailtoassistBrookeldwhentroubleshooting
DV3Toperation.
V.4.1 Log In and Lock Out
TheLogInandLockOutmenuallowstheusertodenetheLogInrequirementandtogglethe
LockOutfeaturewithinRunningViscosityTest(seeFigureV-11).
Figure V-11
TheLogIncheckboxcontrolstheLogInrequirement.WhenLogInischecked,theDV3Twill
requireaUserIDandPasswordpriortoallowinganyactivity.TheLogInscreenispresentedon
poweruporwhentheUserIcon
isselectedfromtheHomeMenu.Theusermustselecttheir
Vedere la pagina 70