Brookeld Engineering Labs., Inc. Page 61 Manual No. M13-2100
Figure V-3
ThecreationofaTemperatureOffsetrequiresinputoftheoffsetvalueandaname.
Theoffsetvaluemustbeintherangeof-9.9to9.9˚C.TheNamecanbeupto14
characterslong.TodeleteanexistingTemperatureOffset,rstselecttheoffsetfrom
thelist,thenpresstheDeletecommandkeyatthebottomofthescreen.
TheuseofaTemperatureOffsetwillbeindicatedintheTemperatureelddisplayed
intheCongureViscosityTestscreenwitha(o)besidetheLiveindication.
TheTemperaturemenualsoallowstheusertoinitiatecontroloveraconnected
temperaturecontroldevicesuchastheBrookeldThermosel(HT-106controller)or
WaterBath(modelTC-XX0SDorTC-XX0AP).SeeSectionII.4forfurtherdetails.
EstablishcontrolovertheexternaltemperaturecontrollerbyselectingTemperature
ControllerintheTemperaturemenu(seeFigureV-4).
Figure V-4
Vedere la pagina 60