Brookeld Engineering Labs., Inc. Page 36 Manual No. M13-2100
AviolationoftheQCLimitsduringthetestwillbeindicatedbyaashingyellowboxaroundthe
displayforthespeciedparameter.NotethatTimedoesnothaveadedicateddisplay.Aviolation
willalsoresultinanaudiblebeepandaonetimewarningmessage(seeFigureIII-11).Datawill
continuetobecollectedwhilethewarningmessageisdisplayed.
Figure III-11
WhenViscosityorTorquearetheselectedQCLimitsparameter,QCLimitswillbeshownduring
thetestontheRealTimeGraphifthatparameterisselectedfordisplay(seeFigureIII-12).The
QCLimitsarerepresentedbythedashedlines.
TIP:AudiblealarmsmaybeturnedoffintheUserSettings
.
TIP: AnindicationofaQCLimitviolationisnotpartofthedataset.Theusercanrecorda
violationbyusingtheNotesavailablewhenviewingtestResults(seeSectionIII.11).