Brookfield LFRA Texture Analyzer Manuale Utente Pagina 11

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
Vedere la pagina 10
Brookeld Engineering Labs., Inc. Page 11 Manual No. M/05-370-B0507
III.4 Probes
TherearearangeofprobesavailableforusewiththeLFRATextureAnalyzer(seeAppendixA).
AllprobesareconguredwitharighthandthreadtoattachtotheLFRA.Attachtheprobewith
care to prevent damage to the threads. Do not overtighten!
The probes may be cleaned while attached to the LFRA. Follow recommendations in Section
I.8.
Do not apply excessive upward, downward or sideways force. Damage may occur to
the load cell.
III.5 Fixtures / Jigs
Therearearangeofxtures/jigsavailableforusewiththeLFRATextureAnalyzer(seeAppen
-
dixA).Thesedevicesaredesignedtoholdasampleinplaceduringmeasurement.Eachxture/
jigisdesignedtoattachtotheBaseTable.Refertotheinstructionsprovidedwiththexture/jig
for proper installation.
Note: Brookeldcandesigncustomxtures/jigstomeetyourapplication.ContactBrook-
eldorourauthorizeddealerfordetails.
III.6 Operating Menu
TheLFRATextureAnalyzeroffersfour(4)testmodesandone(1)calibrationmode.Eachtest
mode requires parameters to be set. These parameters, once set, will be maintained by the LFRA
during a power down to facilitate repetitive testing. The test mode used prior to a power down
will be presented on the screen at the next power up.
All test mode selection and parameter setting is done using the Select/Scroll knob (see Figure
I.2). This knob is both rotated and pressed during normal operation.
Therstlineoftheinstrumentdisplayidentiesthetestmode:Normal, Hold Time, Cycle
Count, Bloom;orcalibrationmode:Static Load (Figure III.3). The user may scroll through
these choices by rotating the Select / Scroll knob (either clockwise or counter clockwise). A test
mode is selected by pressing the Select / Scroll knob.
TEST: NORMAL
TRIGGER: 5.0 (g)
DISTANCE: 3.0 (mm)
SPEED 6.0 (mm/s)
Figure III.3
Vedere la pagina 10
1 2 ... 6 7 8 9 10 11 12 13 14 15 16 ... 45 46

Commenti su questo manuale

Nessun commento