Brookfield DV2T Viscometer Manuale Utente Pagina 67

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
  • Pagina
    / 95
  • Indice
  • SEGNALIBRI
  • Valutato. / 5. Basato su recensioni clienti
Vedere la pagina 66
Brookeld Engineering Labs., Inc. Page 67 Manual No. M13-167-B0614
Viscometer Related Variables
Mostuidviscositiesarefoundtobenon-Newtonian.TheyaredependentonShearRate,time
oftestandthespindlegeometryconditions.ThespecicationsoftheViscometerspindleand
chambergeometrywillaffecttheviscosityreadings.Ifonereadingistakenat2.5RPM,and
asecondat50RPM,thetwocPvaluesproducedwillbedifferentbecausethereadingswere
madeatdifferentshearrates.Thefasterthespindlespeed,thehighertheshearrate.
Theshearrateofagivenmeasurementisdeterminedby:therotationalspeedofthespindle,
thesizeandshapeofthespindle,thesizeandshapeofthecontainerusedandthereforethe
distancebetweenthecontainerwallandthespindlesurface.
Arepeatableviscositytestshouldcontrolorspecifythefollowing:
1) Testtemperature
2) Samplecontainersize(orspindle/chambergeometry)
3) Samplevolume
4) Viscometermodel
5) Spindleused
6) Whetherornottoattachtheguardleg
7) Testspeedorspeeds(ortheshearrate)
8) Lengthoftimeornumberofspindlerevolutionstorecordviscosity
9) Howthesamplewaspreparedand/orloadedintothecontainer
Vedere la pagina 66
1 2 ... 62 63 64 65 66 67 68 69 70 71 72 ... 94 95

Commenti su questo manuale

Sachin Vaijanath Ashture 05 Mar 2025 | 15:48:33

It's good information