- 33 -
The full scale viscosity range for any DV-II+ model and spindle may be calculated using the
equation:
Where:
TK = DV-II+ Torque Constant from Table D2
SMC = Spindle Multiplier Constant from Table D1
The Shear Rate calculation is:
Shear Rate (
1
/
Sec
) = SRC
*
RPM
Where:
SRC = Shear Rate Constant from Table D1
Full Scale Viscosity Range [cP] = TK SMC
10,000
RPM
*
*
Vedere la pagina 32