Brookfield Engineering Laboratories, Inc. Page 45 Manual No. M/92-021-O604
Inaccurate Readings
Verify Spindle, Speed and Model selection
Verify test parameters: temperature, container, volume, method. Refer to:
“More Solutions to Sticky Problems”; Section II.2A - Considerations for Making Mea-
surements
Perform a calibration check. Follow the instructions in Appendix D.
Verify tolerances are calculated correctly.
Verify calibration check procedures were followed exactly.
If the unit is found to be out of tolerance, the unit may be in need of service. See Appendix G for
details on how to return your viscometer.