Brookeld Engineering Laboratories, Inc. Page 63 Manual No. M/07-022-D0613
repeatable value can be obtained without any special effort for any measurement circumstance. But, it
should be known that this type of torque reading will not convert into a correct centipoise value when
using a Brookeld factor if the boundary conditions are not those specied by Brookeld.
The guard leg is a part of the calibration check of the Brookeld LV and RV series Viscometer/Rheometer.
Our customers should be aware of its existence, its purpose and the effect that it may have on data.
With this knowledge, the viscometer user may make modications to the recommended method of
operation to suit their needs.